It is not necessary to switch off the device before cleaning and disinfecting the probe.
Surfaces of the probe must be cleaned and disinfected strictly in the order described below.
1
Point the probe and transducer downwards.
2
Remove any remaining traces of gel with a soft cloth.
3
Clean the surfaces with a soft cloth or a wipe that has been dampened with the recommended cleaning product.
Note: if necessary, wipe down the surfaces with a soft cloth dampened with water.
4
Dry the probe with a soft, dry cloth.
5
Wipe the surfaces using a soft cloth or a wipe soaked in the recommended disinfectant solution.
6
Dry the probe with a soft, dry cloth.
7
Examine the transducer and probe cable for any damage such as cracks, breakage, or liquid leakage.